| My Account | Subscribe | Contact Us | Donate |
You can view the current or previous issues of Diabetes Health online, in their entirety, anytime you want.
Click Here To View
If you are a physician, educator and medical professional who focus on the treatment of diabetes, then this is the must have resource for you.
Finally! A fresh take on the “professional” journal. Each bi-monthly issue cuts through the jargon and presents the most important information you need to enhance your practice and assist your patients.
Each bi-monthly issue of Diabetes Health Professional is a self-contained handbook covering products, educational resources and the latest diabetes research, complimented by balanced editorial focused on medical news, drug prescription information, clinical practice recommendations and changing treatment options.
Each quarter we send you the latest, most updated research guides, product guides and educational resource guides available for you and your patients.
Each week the Diabetes Health E-Newsletter delivers links to the very latest in news, reviews, blogs and videos from Diabetes Health direct to your inbox.
As a subscriber you'll get access to the amazing Diabetes Health Digital Advantage™ so you can read the current issue of Diabetes Health magazine online wherever you are!
You can cancel your newsletter subscription at anytime by clicking "Unsubscribe" on the bottom of any newsletter you receive
Then enter your new email address in the above form and click "Subscribe"
Latest Professional Issues Articles
The infusion sets in question have the reference numbers MMT-396, MMT-397, MMT-398, and MMT-399, and lot numbers starting with the number 8 printed on their packaging.
Minneapolis-based Medtronic Inc. is recalling some lots of its Quick-set infusion sets over concerns that they may cause insulin pumps to deliver too much or too little insulin.
The infusion sets in question have the reference numbers MMT-396, MMT-397, MMT-398, and MMT-399, and lot numbers starting with the number 8 printed on their packaging.
Customers are asked to return the affected infusion sets to the company. For more information, consumers can call 800-345-8139 or visit http://www.medtronicdiabetes.com/lot8.
Users experiencing adverse reactions can report them to the Food and Drug Administration at http://www.fda.gov/medwatch
The infusion sets, used with the company's MiniMed Paradigm insulin pumps, deliver insulin via a thin plastic tube from the pump to the patient. A malfunction could cause them to deliver too high or to low a dose of insulin, leading to hyperglycemia or hypoglycemia-both dangerous conditions for people with diabetes.
However, Medtronic says that there have been no reports of any malfunctions among users of the infusion sets. The recall is a precautionary measure intended to thwart a potential problem.
(Editor's Note: The message below, distributed by Medtronic, explains the company's recall decision. We are publishing it unaltered to assist Medtronic in getting the word out to Quick-set users.)
A Message from Medtronic
To Customers Previously Using "Lot 8" Quick-set* Infusion Sets
Please know that we care about your safety - that's why Medtronic acted quickly to recall "Lot 8" Quick-set infusion sets. We would like to share some background information about this recall, and also describe how clogged vents on an infusion set could affect blood sugar levels.
Background Information
Please know that Medtronic voluntarily initiated this recall - it was not triggered by a patient injury report. Medtronic received a report that an infusion set did not prime properly and our internal testing revealed that this was a result of clogged vents due to a change in the manufacturing process. Venting is necessary to equalize air pressure in the reservoir compartment with the surrounding atmosphere. Clogged vents could potentially cause too much or too little insulin to be delivered.
The manufacturing process was corrected and Medtronic is instructing patients to stop using "Lot 8" Quick-set infusion sets. Last week, Medtronic sent more than one million infusion sets to customers to ensure therapy continuation. The company is also working closely with distributor partners to provide infusion sets to everyone who needs them.
Blood Glucose Variability and Clogged Vents
Although patients may wonder if their past blood sugar variability is a result of clogged vents, our internal testing shows that the possibility for insulin over or under delivery is limited to the following two scenarios:
Scenario 1:
A significant over-delivery of insulin caused by a rapid increase in altitude, potentially leading to a severe hypoglycemic event.
Scenario 2:
The non-delivery of insulin closely following an infusion set change, potentially leading to a hyperglycemic event.
Scenario 1-Over-Delivery of Insulin
For a significant hypoglycemic event to occur as a result of clogged vents, a rapid change in air pressure would be necessary. For example, when taking off in an airplane or quickly driving up a mountain, air pressure could build up inside the reservoir compartment and potentially cause over-delivery of insulin. This could result in a severe hypoglycemic event, which may require immediate action or assistance.
Since our testing shows that this occurs only with a rapid increase in altitude, this severe low blood sugar event is different than the typical low blood sugars that many people experience as part of their routine diabetes management.
Scenario 2-Non-Delivery of Insulin
During priming, clogged vents may prevent the reservoir compartment from equalizing air pressure. As a result, the insulin pump may appear to be properly primed when it is not. Once the infusion set is inserted, this could lead to a period of non-delivery of insulin, which can potentially result in a hyperglycemic event following an infusion set change.
Since this occurs closely following an infusion set change, it may be different than the typical high blood sugars that many people experience as part of their routine diabetes management
So that we can serve all of our customers during this recall, we will be sending infusion sets at a rate of approximately one box every three weeks - we are doing this to ensure therapy continuation for everyone. We deeply apologize for the inconvenience of this process. Please know that patient safety is our highest priority.
-Medtronic Diabetes