LifeScan Removes a Single Lot of Test Strips

3916

In a January 20, 2004 letter, LifeScan informed its OneTouch Ultra Test Strip customers it was voluntarily removing a single lot of 50-Count OneTouch Ultra Test Strips.

The lot number of the affected test strips is #1022859.

“LifeScan has determined that the Control Solution Range was printed incorrectly on the vial label of one lot of 50-count OneTouch Ultra Test Strips sold in the United States,” wrote LifeScan. “The control solution range for these strips was incorrectly printed as 97-131 mg/dL (5.4-7.3 mmol/L) instead of the correct range of 111-150 mg/dL (6.2-8.3 mmol/L).”

No other OneTouch Ultra Test Strips or any other OneTouch products are affected by this action.

If you have any questions, call LifeScan’s toll-free 24-hour customer service line at (888) 849-7386.

* * *

Source: LifeScan, January 20, 2004 press release

Click here to view/write comments
Diabetes Health Medical Disclaimer
The information on this site is not intended or implied to be a substitute for professional medical advice, diagnosis or treatment. All content, including text, graphics, images, and information, contained on or available through this website is for general information purposes only. Opinions expressed here are the opinions of writers, contributors, and commentators, and are not necessarily those of Diabetes Health. Never disregard professional medical advice or delay seeking medical treatment because of something you have read on or accessed through this website.